Horizontal Rectangular Pressure Steam Sterilizer

Product Details
Customization: Available
Theory: Pressure Steam
Type: Pressure Steam Sterilization Equipments
Manufacturer/Factory & Trading Company
Diamond Member Since 2020

Suppliers with verified business licenses

Audited Supplier

Audited by an independent third-party inspection agency

to see all verified strength labels (15)
Find Similar Products

Basic Info.

Model NO.
YXQ. WF21-0.2/0.36/0.4/0.5/0.6/0.8SZ
Certification
CE
Electric Tension
380V
LCD Display
Without LCD Display
Sterilization Information Recording and Printing
Without Sterilization Information Recording and Printing
Material
Stainless Steel
Power
AC.380/220±10%(V)50±1(Hz)
Working Pressure
0.14
Working Temperature
126ºC
Transport Package
Standard Export Package
Trademark
Labkits
Origin
Hunan China
Production Capacity
500PCS/Month

Product Description

Item: Horizontal Rectangular Pressure Steam Sterilizer
Model: YXQ.WF21-0.2/0.36/0.4/0.5/0.6/0.8SZ
Horizontal Rectangular Pressure Steam Sterilizer
Characteristics:

1. Main body adopts rectangular double-layer structure. Inner layer is high quality 06Cr19Ni10 stainless steel and outer layer is pressure vessel steel plate Q235-B. Both layers are welded by adopting automatic welding technology. Inner layer is treated with special mechanical process and chemical surface technology. 
2. Safe reliable full-automatic program operation control over pressure, temperature and time, LED display screen, dialogue frame of displaying whole sterilization process. Sterilization process is controlled automatically to ensure easy operation, without need of monitoring.
3. Sterilization temperature range 105-1290C is available, sterilization procedure can be set freely to sterilize different types of items.
4. This sterilizer automatically starts countdown function when entering drying procedure, automatically cut off heating power after drying is finished, and gives alarm to indicate procedure operation is completed.
5. It is equipped with over-temperature, overpressure automatic protective device.
6. When the sterilizer door is not closed tightly, steam cannot enter inner chamber body, then give alarm, the door is interlocked and cannot be opened when the pressure inside sterilization chamber is more than 0.027Mpa.
7. Safety valve automatically release pressure for protection when steam pressure exceeds operating pressure.
8. Heating cannot be realized when level is high, and protective device automatically cuts off power and gives alarm when water level is low.
9. Power breaker automatically cuts off power in time of current over.
10. Air-tight door adopts radial latch locking structure, and is equipped with safe interlocking device to ensure easy operation. Sealing ring adopts high temperature resistant, highly elastic silicon rubber to ensure long durability.
 
Technical Parameters:
1. Design pressure (MPa): 0.17
2. Design temperature (ºC): 130
3. Working pressure (MPa): 0.14
4. Operating temperature (ºC): 126
5. Water pressure (MPa): 0.15-0.3 (deionized water)
Volume
(m3)
Inner chamber dimension (L*W*H) mm Water consumption
<(kg)
Stroke cycle
<(min)
Weight
(KG)
Control power
<(kw)
Electric source Electric steam power
≤(kw)
0.2 750×500×550 30 80 900 0.5 Three-phase four-wire system AC.380/220±10%(V)50±1(HZ) 18
0.36 1000×580×620 30 80 1050 0.5 24
0.4 1150×580×620 40 80 1100 0.5 24
0.5 1000×680×720 40 80 1150 0.5 24
0.6 1200×680×720 40 80 1200 0.5 24
0.8 1500×680×720 50 80 1300 0.5 36