High and Low Temperature Vacuum Friction and Wear Testing Machine

Type: Friction Testing Machine
Maxcapacity: <1000KN
Accuracy Grade: 1
Load Way: Electronic Load
Loading Method: Dynamic Load
Display: Digital
Customization:
Manufacturer/Factory & Trading Company
Diamond Member Since 2020

Suppliers with verified business licenses

Rating: 5.0/5
Hunan, China
to see all verified strength labels (15)

Basic Info.

Model NO.
MDZ-05GL
Control
Computer Control
Weight
0-100Kg
Power Source
AC220V
Axial Test Force Working Range
0.1-10n 10-500n
Max Friction Torque
2500n.mm
Spindle Speed Range
5~2000r/Min
Vacuum
10-2PA
Low Temperature Range
-160 ° C ~ Room Temperature
Temperature Control Accuracy
± 5 ° C
Transport Package
Standard Export Package
Trademark
Labkits
Origin
Hunan China
Production Capacity
500PCS/Month

Product Description

Item: High and low temperature vacuum friction and wear testing Machine 
Model: MDZ-05GL
High and Low Temperature Vacuum Friction and Wear Testing Machine
 

Application

The high-temperature vacuum friction and wear tester is under a certain pressure and speed, under high temperature, low temperature, vacuum and special gas protection environment, simulates the actual working conditions, and controls and changes the load, speed, time and other parameters to test. Evaluate the comprehensive friction and wear performance of materials such as friction coefficient, wear amount, and depth of wear scars.
 

Technical parameters

No

Name

Technical indicators

1

Test force

 

1.1

 

Axial test force working range

0.1-10N

10-500N

1.2

Relative error of test force indication

≤±1%

1.3

Relative repeatability error of test force indication

≤1%

1.4

Loading and unloading test force should be stable, without impact and pulsation

 

1.5

Error of displayed value change during long test force

≤±1%FS

1.6

Test force display resolution

0.1-10N / 0.1N

10N-500N / 1N

2

Frictional force

 

2.1

Max Friction torque

2500N.mm

2.2

Relative error of friction torque indication

≤±2%

2.3

Repeated relative error of friction torque indication

≤2%

2.4

Friction torque display resolution

0.1N/1N.mm

3

Spindle speed

 

3.1

Spindle speed range

5~2000r/min

3.2

Above 100r / min error for spindle speed range

≤±5r/min

3.3

100r / min and below 100r / min error

≤±1r/min

4

Vacuum

10-2pa

5.1

Low temperature range

-160 ° C ~ room temperature

5.2

Temperature control accuracy

± 5 ° C

5.3

High temperature range

Room temperature ~ 1000 ° C

5.4

Temperature control accuracy

± 2 ° C

6

Testing machine time display and control range

1~99999s or min

7

Testing Machine revolution display and control range

999999999

Send your message to this supplier

*From:
*To:
*Message:

Enter between 20 to 4,000 characters.

This is not what you are looking for? Post a Sourcing Request Now